TM 10-6640-215-13 4-47.  REMOVE/INSTALL THE DISTILLATION TEST APPARATUS- continued. 2. Remove Condenser Assembly.  See figure 4-47. a. Remove shield assembly as in paragraph 1. above. b. Unplug the condenser power cord from the wall receptacle. c. Remove drain lines(s). d. Remove one screw from connecting mounting bracket. e. Remove six screws from mounting bracket. f. Remove defective condenser. INSTALLATION 1. Install Condenser Assembly.  See figure 4-47. a. Using mounting plate as template, scribe and drill holes in new condenser. b. Place mounting bracket on new condenser and secure with five mounting screws. c. Connect drain line(s). d. Install six condenser assembly mounting screws attaching assembly to the countertop. e. Install one screw in connecting bracket. f. Plug in unit and verify operation. 2. Install Shield Assembly.  See figure 4-47. a. Using mounting plate as template, scribe and drill holes in new shield assembly. b. Place mounting bracket on new shield assembly. c. Install five shield assembly mounting screws. d. Secure shield assembly to countertop with mounting screw. e. Connect shield to condenser. 4-128


 